Covidien Capnostream 20p Capnograph
Covidien

Covidien Capnostream 20p Capnograph

Covidien | SKU
20p1
In Stock
$1095.00
Select Accessory

The Capnostream 20p patient monitor is built on a legacy of proven performance. For nearly two decades, clinicians have relied on Microstream® technology for an accurate, continuous view of ventilation adequacy on intubated and non-intubated patients, from neonate to adult. The new Capnostream 20p monitor, with the Apnea-Sat Alert feature, alerts caregivers to repetitive patterns of apneas and oxygen desaturation, revealing a more complete picture of a patient’s respiratory status.

Capnostream 20p Brochure
Download Document of Covidien Capnostream 20p Capnograph.
Discount Cardiology ships via the Federal Express Service, which delivers shipments anywhere in the United States -- usually within 2-5 business days -- at low rates. Unless specified, shipping charges are added to each order based on the specific order.

Covidien Capnostream 20p Capnograph

This is a refurbished Capnograph that monitors ETCO2 and SPO2. It will arrive "patient ready" with a 1yr warranty.

Capnostream 20p Bedside Patient Monitor Features

Dual parameter monitor

  • - Supports current standards of care by providing etCO2 and SpO2 measurements.

Smart Alarm Management Integrated Algorithms 1-2

- The Capnostream 20p patient monitor with Microstream includes Smart Alarm Management Integrated Algorithms, a suite of algorithms proven to reduce alarms and simplify the use of patient monitoring to enhance patient safety and improve clinical efficiency.

Smart Breath Detection

  • - Screens out low amplitude ‘non-breath’ etCO2 excursions like, talking, crying or snoring.

Smart Alarm for Respiratory Analysis (SARA)

  • - Reduces clinically insignificant alarms.

Nellcor™ SatSeconds

  • - Alerts clinicians to clinically significant changes in pulse oximetry.

Integrated Pulmonary Index Algorithm (IPI )

  • - IPI algorithm presents one value that demonstrates real-time respiratory status based on etCO2, RR, PR and SpO2.††

Apnea-Sat Alert Algorithm (ASA ) 

  • - Tracks and reports apneas per hour and oxygen desaturation index, and provides the data in trend reports.

Connectivity

  • - Compatible with central monitoring systems and interfaces with nurse call systems and electronic medical records.

Internal printer

  • - Provides convenient bedside documentation.

SARA evaluates the stability of the detected breath-to-breath intervals and adjusts (adapts) the measurement period accordingly, thus averaging out noise and transient fluctuations when required, providing a more realistic respiratory rate.

††SpO2 data from host required.

†††Apnea-Sat Alert is FDA cleared for adults 22 years of age or older.

1. ECRI Institute. The Hazards of Alarm Overload: Keeping Excessive Physiologic Monitoring Alarms from Impeding Care. ECRI Guidance Article, March 2007.

2. Hockman S, Glembot T, Niebel K. Comparison of capnography derived respiratory rate alarm frequency using the SARA algorithm versus an established nonadaptive respiratory rate alarm management algorithm in bariatric surgical patients.Resp Care (Open Forum Abstracts). 2009;12

Capnostream 20p Bedside Patient Monitor Specifications

Specification Overview:

General characteristics

Unit dimensions:

167 mm. (h) x 220 mm. (w) x 192 mm. (d)

 

(6.6 in. (h) x 8.7 in. (w) x 7.6 in. (d))

Unit weight:

3.6 kg.

 

7.94 lb.

Packaged dimensions:

315 mm. (h) x 340 mm. (w) x 285 mm. (d)

 

12.4 in. (h) x 13.4 in. (w) x 11.2 in. (d)

 

Power supply

Input Voltage 100-240 V, 50/60 Hz, ± 10%.

 

Microstream® capnography

• Units: mmHg, kPa or Vol%.

• Measurement range etCO2, FiCO2 0-150 mmHg.

• Respiratory Rate 0 bpm-150 bpm.

• Resolution 1 mmHg.

• Accuracy (±SD) etCO2, FiCO2 0-38 mmHg ± 2 mmHg. 39-150 mmHg ± (5% of reading + 0.08% for every 1 mmHg above 38 mmHg).

• Respiratory rate: 0-70 bpm ±1 bpm, 71-120 bpm ±2 bpm, 121-150 bpm ±3 bpm.

• Flow rate 50 ml/minute, flow measured by volume.

• Waveform sampling 50 millisecond intervals.

• Response time 2.95 second (typical).

• Initialization time 40 second (typical).

• Calibration interval Initial calibration after 1,200 operating hours, then annually or after 4,000 operating hours, whichever comes first.

 

Nellcor pulse oximetry

• Measurement range 0-100%.

 

• Accuracy

Adult mode:

 

o SpO2 range 70%-100% ± 2 digits.

 

o SpO2 range 0-69% unspecified.

 

Neonate mode:

 

o SpO2 range 70%-100% ± 3 digits.

 

o SpO2 range 0-69% unspecified.

• Pulse rate

o Range: 20-250 BPM.

 

o Accuracy: ± 3 BPM.

 

o Sat Sec Range: 10, 25, 50, 100.

 

There have been no reviews