Datascope Passport EL Multi-Parameter Patient Monitor
Datascope

Datascope Passport EL Multi-Parameter Patient Monitor

Datascope | SKU
Datascope Passport1
In Stock
Call For Pricing : 1-800-507-8244
Select Accessory

Datascope Passport EL Multi-Parameter Patient Monitor at the lowest price. Call for details : 1-800-507-8244

Discount Cardiology ships via the Federal Express Service, which delivers shipments anywhere in the United States -- usually within 2-5 business days -- at low rates. Unless specified, shipping charges are added to each order based on the specific order.

Product Description

The Datascope Passport monitor measures ECG, Respiration, NIBP, SPO2, and Temperature.  The Passport has a large electroluminescent screen for easy viewing from any angle. This monitor is operated with easy to use, single function keys.

Specifications:

  • Dimension: 9.5" x 13" x 8" (H x W x L)
  • Weight: 13 lbs. maximum without batteries
  • Power Consumption: 15 watts typical
  • Voltage input: 100, 120, 220, 240 VAC ± 10%
  • Mains frequency: 47 to 65 Hz 

ECG

  • Leads: I, II, III
  • Display sensitivity: 0.25, 0.5, 1.0, 2.0, 3.0, 4.0 cm/mV ± 10%
  • Frequency response to screen: ESU filter out: 0.5 to 40 Hz, -3db max @ 1 mVpp input; ESU filter in: 0.5 to 20 Hz -3db max @ 1 mVpp input

Heart Rate Meter

  • Range: 30-300 bpm adult/child. 30-350 bpm neonatal
  • Accuracy: 30-250 bpm ± 3bpm or 3%, whichever is greater.
  • Pacer rejection: rejects all pulse of amplitude ± 2.0 mV to ± 700 mV and duration 0.1 ms to 2 ms with no tail. Rejects all pulses ± 2.0 mV to ± 700 mV and duration 0.1 ms to 2 ms with 100 ms T.C. tail <0.8 mV or 25 ms T.C. tail <0.4 mV.
  • Display update: 2 seconds ± 0.1 sec

NIBP

  • Technique: oscillometric
  • Systolic range: adult/ped mode: 40-250 mmHg, neonate mode: 15-150 mmHg.
  • Diastolic range: adult/ped mode: 30-200 mmHg, neonate mode: 10-150 mmHg.
  • Systolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric). Mean error less than 3 mmHg, standard deviation less than 13 mmHg (neonatal).
  • Diastolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric). Mean error less than 4 mmHg, standard deviation less than 7 mmHg (neonatal).
  • Heart rate meter range: 30-250 bpm.
  • Heart rate meter accuracy: ±3 bpm or ±3%, whichever is greater.

Pulse Oximetry

  • Accuracy (1standard deviation): 70-100%, ±2%; 60-69%, ±4%; 0-59%, unspecified.
  • Response time: 13 seconds to 95% of final step change % SpO2 value from 60-95% at 72 bpm

Recorder (optional)

  • Speed: 50, 25, 12.5, 6.25, 3.125 mm/sec ±5%
  • Note: 6.25 and 3.125 are for respiration only
  • Chart size: 50 mm ±2mm

Battery Requirements

  • 2 sealed lead batteries
  • Voltage: 12 VDC
  • Capacity: 2.3 AH each
  • Run time: 2 hours fully charged battery
  • Recharge time: 16 hours maximum
  • Low battery warning: 11.6 ± 3.5%
  • Low battery cut off: 11.0 ± 2.8%

Additional Information

  • Dual-Vector ECG, Respiration, NIBP, SpO2, 2 Temperatures, and 2 Invasive Pressures in one compact SuperModule.
  • Multi-Lead ST Analysis and Arrhythmia Detection with waveform recall and analysis also standard.
  • Optional Cardiac Output module with hemodynamic calculations and output averaging cost-effectively expands capabilities.
  • Optional Dual Invasive Pressure modules add up to four additional invasive pressures.
  • High intensity, high resolution 10.4" removable color LCD display mounts easily to anesthetic gas machine.
  • Touchscreen and dedicated keys ensure fast, direct operations 6 user-programmable soft keys, 3-format display memory, and choice of 32 parameter colors provide maximum flexibility.
  • PCMCIA data card allows patient data and customized set-ups to be saved or transferred to another Expert Simultaneous display of up to 6 waveforms and 100-waveform event storage facilitate patient management.
  • Main unit, module housing, and display can be separately mounted when space is at a premium.
  • Optional Datascope Gas Module II provides breath-by-breath Fast O2, CO2, and N2O monitoring with automatic or manually-controlled 5-agent anesthetic measurement for real-time assessment
There have been no reviews


: *
: *
: *