QRS Orbit Portable Spirometer
The Orbit is a full-function portable spirometer which connects directly to your Personal Computer, laptop or tablet. This spirometer simplifies testing by eliminating calibration and sterilization with disposable nose-clips and pre-calibrated mouthpieces. The Orbit integrates with Office Medic and cooperates with your EMR system.
Measured Parameters – FVC, FEV0.5, FEV1, FEV6, FEV3, FEV1/FEV6, FEV1/FVC, FEV3/FVC, PEFR, PEFT, FEF25%, FEF50%, FEF75%, FEF25-75%, FIVC, FIV0.5, FIV1, FIV3, FIV1/FIVC, FIV3/FIVC, PIFR, FIF50%, FIF25-75%, FIF.2-1.2, FVC/FIVC, Extrapolated Volume, EOTV, FET, MVV, RR, MTV, SVC
Language Options – English, Spanish, French, German, Japanese, Italian, Portuguese
Standards – The Orbit provides effective diagnostic tests and meets the following Standards:
- American Thoracic Society (ATS/ERS) 2005 Standard
- British Thoracic Society (BTS) Standard
- National Institute for Health & Clinical Excellence (NICE) Guidelines
Lung Age Calculation – Lung age is calculated for patients ages 20-84. It is equal to the predicted FEV1 that matches the patient’s actual FEV1. Lung age may differ based on the predicted equation chosen.
Patented, Pre-Calibrated Mouthpiece – With the Orbit spirometer system, everything the patient touches is thrown away. No sterilization or calibration is necessary, saving time for nurses, therapists and physicians.
No Batteries Required – The Orbit draws its power from the PC it is connected to by USB port.
The Orbit comes equipped with everything you need to begin testing.
- Orbit Portable Spirometer
- QRS Pre-Calibrated Mouthpieces, 2
- Pressure Tubes, 2
- Nose Clips, 5
- Office Medic Software
Also included with the Orbit portable spirometer is a three-year warranty.
Weight: | 0.5 lb. |
Height: | 4.3” x 3.7” x 1.7” |
Range (BTPS): |
FLOW: ±14 liters/second VOLUME: 0.5-8 liters |
Accuracy (BTPS): |
FLOW: ±5% of indication or ±200 ml/sec, whichever is greater for FEF 25-75 and ±10% of indication or ±300 ml/s whichever is greater for PEF VOLUME: ±3% of indication or ±50 ml, whichever is greater for FVC and FEV1; ±10% of indication or ±15 L/min, whichever is greater for MVV |
Precision (BTPS): |
FLOW: 5% of indication or 150 ml/sec, whichever is greater for PEF VOLUME: 3% of indication or 50 ml, whichever is greater for FVC and FEV1 |
Calibration: | ATS 3-speed or standard calibration check |
Predicted Normals: |
Adult: Crapo, Cherniack, Morris, Knudson 1983, Roberts, NHANES III, ECC/ERS/Quanjer Pediatric: HSU, Knudson 1983, Polgar, Warwick, NHANES III, Zapletal, Wang, Quanjer |
Tests Performed: | FVC, Pre/Post Testing, Flow Volume Loop, MVV, SVC |
Measuring Time: |
FVC – 60 seconds SVC – 60 seconds MVV – 15 seconds |
Power Supply: | 5 Vdc ±5% 100 mA from the host PC USB Port |
Operating Conditions: |
Temperature: 15-40°C Humidity: 10-90% (non-condensing) Atmospheric Pressure: 700-1060 hPa |
Storage Conditions: |
Temperature: -15-50°C Humidity: 90% max. (non-condensing) Atmospheric Pressure: 700-1060 hPa |