Description
NELLCOR N600X
Â
The Nellcor N-600x pulse oximeter with OxiMax technology delivers exceptional performance throughout the hospital, so clinicians can confidently monitor even their most critically ill patients.
While some oximeters may have trouble acquiring a pulse signal during low perfusion and signal interference, the Nellcor N-600x pulse oximeter with OxiMax technology achieves accurate, reliable readings during these difficult conditions.
The Nellcor N-600x pulse oximeter includes our unique SatSeconds alarm management technology to safely reduce nuisance alarms that can interfere with staff productivity
Â
Â
·                    Saturation Pattern Detection (SPD) alert
·                    Real-time remote monitoring capabilities
·                    Digital signal processing technology that enhances performance during difficult conditionsâ€
·                    SatSeconds alarm management function for a safe, practical way to reduce nuisance alarms3
·                    Connects to the Covidien remote respiratory monitoring systems or nurse call systems for enhanced patient safety on general care floors
Â
Exceptional performance in any clinical environment and compatibility with the complete line of Nellcorâ„¢ sensors with OxiMaxâ„¢ technology.
To help patients thrive, clinicians need accurate and timely patient oxygenation information. With a complete portfolio of monitoring technologies, the sensing systems of Covidien enable caregivers to detect subtle, yet critical variations in patient status for faster, more informed interventions.
·        Offers a convenient selection of adult or neonate alarm limits and user-configurable power-on settings
·        Includes an on-screen quick guide for instant help to assist caregivers with monitor operation
·        Features a sensor messages function that provides troubleshooting tips for optimal sensor application
·        Uses a variable pitch beep tone to enable clinicians to hear point-by-point changes in SpO2
·        Offers user-selectable fast-averaging mode (two to three seconds) to more precisely track rapidly changing saturations
·        Provides on-screen viewing of 24 to 48 hours of downloadable SpO2 and pulse rate trends taken at two- to four-second intervals, with printing capability
·        Compatible with Philips VueLink* device interface module, M1032A-A05
Â
Â
PERFORMANCE
| Â | Â |
| Measurement Range | Â |
| SpO2: | 1% to 100% |
| Pulse rate: | 20 to 250 beats per minute |
| Perfusion range: | 0.03% to 20% |
ACCURACY1
| Saturation : | (% SpO2 ± 1 SD) |
| 70% to 100% : | ± 2 digits |
| 60% to 80% : | ± 3 digits |
| Low perfusion: | 70% to 100% ± 2 digits |
| Pulse rate: | 20 to 250 bpm ± 3 digits |
| Low perfusion : | 20 to 250 bpm ± 3 digits |
ELECTRICAL
| Instrument | Â |
| Power requirements: | 100 to 120 VAC, 200 to 240 VAC, 50/60 Hz, 20 VA |
| Fuse rating: | 0.5 amperes |
Â
| Battery | Â |
| Type: | Sealed lead-acid internal battery |
| Battery capacity: | Typically seven hours using new, fully charged battery |
| Charge time: | 8-12 hours |
PHYSICAL CHARACTERISTICS
| Weight |
| 5.7 lb (2.6 kg) |
Â
| Size |
| 3.3 in. H x 10.4 in. W x 6.8 in. D (8.4 cm x 26.4 cm x 17.3 cm) |
Â
| Equipment Classification |
| IEC 60601-1 |
| EN865 |
| CSA 601.1 |
| IPX1 |
| UL 2601-1 |
| ISO 9919 |
| EN/IEC 60601-1-2 (2nd ed.) |
Â
| Emissions Compliance |
| EN55011, Group 1, Class B |
Â
| Output |
| EIA-232 and RS-422 digital output from serial port |
| Nellcorâ„¢ OxiNet Communications Output Protocol |
| Nurse call capability |
| Analog output (0-1V); SpO2, pulse rate and plethysmographic waveform |
Â
| Displays/Indicators |
| High-quality LCD graphical screen |
| Visual indicators: Pulse search, audible alarms silenced or off, interference indicator, low battery and battery charging, battery fuel gauge, sensor off, sensor disconnect, data-in sensor and SatSeconds alarm management clock |
| Â |
Â
| Alarms |
| Audible and visual alarms for high/low saturation and pulse rate |
| SatSeconds alarm management settings: 10, 25, 50 and 100, or off |
| Audible and visual warning indicators for low battery |
| Audible and sensor disconnect alaram and sensor-off alarm |
| Displayed saturation and pulse rate alarm limits |
| Backup speaker with escalating alarm tones |
Â
| Optional Accessories |
| GCX mounting options |
| Interface cables |
| Carrying cases |
Â
Â