Description
Edan SE -601a Features:
- FDA Approved
- Heart rate variability (HRV) & R to R Trending analysis
- Portable, lightweight design
- Easy data input and operation
- Alphanumeric keyboard and one-touch operation
- Built-in rechargeable battery, AC/DC power supply
- Pacemaker detection which meet the requirements of ANSI/AAMI EC11
- Automatic analysis and diagnostic software for adults and pediatrics
- Two-step exercise test with periodic recording
- Internal thermal printer and external printer
- Support external archiving: USB flash disk, card reader
- Data transmission to PC via Ethernet or serial port
- 2 year extended warranty available for $ 299
| Model | Screen | Internal speaker | Interpretation | Internal storage |
| SE-601A | 3.5 inch STN LCD, 192*64 dots | buzzer | Yes | 50 ECGs |
| SE-601B | 5.7 inch STN LCD, 320*240 dots | loudspeaker | Yes | 100 ECGs |
| SE-601C | 5.7 inch TFT color LCD, 640*480 dots | loudspeaker | Yes | 100 ECGs |
| Optional Functions | XML format export, Internal storage upgrade to 200 ECGs, Real-time data transfer to PC, GSM/GPRS wireless data transfer, External standard keyboard | |||
Technical Specifications:
Â
- Input Circuit: Defibrillator protection
- Lead: Standard 12 leads
- Acquisition Mode: 12 channel simultaneous acquisition (10-24s adjustable)
- 6 Lead waveform previewÂ
- stores 100 EKG's internally
- A/D Converter: 24 bits
- Sampling rate: 1000Hz/channel, 10,000HZ for pacemaker detection
- Noise Level: ±5mVp-p
- Time Constant: >3.2s
- Frequency Response: 0.05Hz - 150Hz
- Calibration Voltage: 1mV±2%
- Sensitivity: 2.5, 5, 10, 20, 10/5, AGC (mm/mV)
- Input Impedance: >50MΩ
- Input Circuit Current: <10μA
- Noise Level: <12.5μVp-p
- Patient Leakage Current: <10μVp-p (220V-240V)
- Patient Auxiliary Current: <0.1μA (DC)
- Filter:Â
EMG Filter: 25Hz/35Hz/45Hz/OFFÂ
Anti-baseline drift: 0.05/0.15/0.25/0.5HzÂ
Low-pass Filter: 75Hz/100Hz/150HzÂ
AC Filter: 50Hz/60Hz - CMRR: >115dB
- Recording Mode: Auto/Manual/Rhythm
- Rhythm lead: Single channel or three channels selectable
- Power Supply:Â
AC: 100-240V, 50/60HzÂ
12V Vehicle-Carried Inverter
Li-on battery - Battery: Built-in rechargeable Li-on battery, 2200mAh, 14.8V for 6 hours general use or 280 copies ECGs continuous printing
- Recorder: Thermal printer
- Record Paper: 110mm x 140mm x 144p (Z-fold)
- Paper Speed: 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s (±3%)
- Weight: Approx. 3kg (including battery and paper)
- Dimension: 322mm x 310mm x 101mm
- Display:Â
5.7 inch B & W STN LCD, 320x240 dots (SE-601B)Â
- External Input/Output:Â
Input: > 100kΩ ; Sensitivity 10mm/V ±5%Â
Output: < 100Ω ; Sensitivity 1V/mV ±5% - Communication Interface: Ethernet, RS232